Beam Behaviour and Magnetic Field of K500 Scc

Vinay Singh#, A Dutta, A Agarwal, C Das, B Naik, Z A Naser, S Paul, U Bhunia, J Pradhan, J Debnath Variable Energy Cyclotron Centre, 1/AF Bidhan Nagar, Kolkata 700064
Abstract
The extraction of the beam from the K500 Superconducting Cyclotron in Kolkata is posing a great challenge. After getting the internal beam, lots of effort was made to extract the beam and thereafter various experiments were designed and done to know the beam behaviour and to investigate the real reason. The deflector electrode was used in the Faraday cup mode. The deflector probe, bore scope probe and the main probe were used to investigate the beam behaviour by varying the first harmonic amplitude and phase with the help of trim-coil no.13 operated in harmonic mode. Finally magnetic field mapping was done. In this paper, all these experimental results are described in detail.
INTRODUCTION
K500 SCC got its first internal beam till the extraction radius in Aug 2009 [1]. This was confirmed by presence of neutrons in the neutron detector. The main probe and zinc sulphide screen (bore scope) were the only diagnostics available then. After installing the deflector, numerous runs were tried but the beam couldn’t be guided through it. A series of experiments were designed to get into the bottom of the problem. It was then found that the magnetic field was the main reason behind this. This was further confirmed by magnetic field mapping . In this paper the steps leading to this conclusion are described in detail.
The electrostatic deflector
Deflector as a beam diagnostic element
In K500 SCC there are two sets of deflector E1 and E2. The beam first encounters the deflector E1 which spans from 335° to 34°. In the normal mode i.e., when the power supply is connected to it, and electric field of around 2.5 MV/m is created by applying high voltage of 20kV between the electrodes. The gap between them is 8mm.
To eliminate the possibility, the beam was at all going through the deflector or not, the power supply was disconnected and the deflector was operated in Faraday mode i.e., as a beam diagnostic element.
The deflector E1 is not a fixed element, but has certain amount of freedom in respect to its radial distance from the centre of cyclotron. The entry position can be varied from 667 to 671 mm and the exit from 672 to 678 mm.
The variation in position of deflector coupled with use of the 1st harmonic coils (to slightly alter the beam curvature) was done repeatedly to get the beam signature on the outer electrode. However no significant beam current could be detected.
Use of wire mesh at the entry and copper block at the exit of deflector
To further investigate a wire mesh was placed at the entry of the deflector (Fig. 1) and a solid copper block at the exit (Fig. 2). This was done to ensure that any trace of beam reaching the deflector entry can be detected. And this beam if successfully passes through the deflector will be detected by the copper block. So, this time the deflector was used in the normal mode i.e., in the power supply mode.

Don't use plagiarized sources. Get Your Custom Essay on
Beam Behaviour and Magnetic Field of K500 Scc
Just from $13/Page
Order Essay

Get Help With Your Essay
If you need assistance with writing your essay, our professional essay writing service is here to help!
Essay Writing Service

Wide range of 1st harmonic magnetic field amplitude and phase with the help of trim coil no.13 was applied. This was done to vary the beam trajectory near the extraction zone so as to get beam signal from the entry mesh. However no significant beam current could be detected. To further explore this point, the deflector voltage and its orientation were changed through a series of iterative methods, but no current could be detected on the exit copper block.
————————-
#vsingh@vecc.gov.in

 

Figure 1: View of wire messed current diagnostic at the entry of Deflector

Figure 2: Solid copper block current diagnostic at the exit of Deflector

Use of polymer sheets for Beam impression
To get a further insight, another attempt was made by putting polymer sheets on the inner portion of the septum at three different locations – one near the entry , second at middle and the last near the exit. After a small amount of beam time the sheets were taken out for chemical analysis. This was done to exclude the possibility of any marks that could have arisen due to the high voltage discharge of the electrode. Only a faint beam impression (Fig. 3) was found on the sheet placed near the deflector and no impression was found on the other two placed further away.

Figure 3: Beam mark on plastic sheet placed on deflector

Beam characteristics using probes
The beam appeared to be not reaching the deflector side, but at the same time the beam was very much available towards the main probe and bore probe, so further investigations were continued to know the off-centeredness of the beam.
Description of the Probes
The K500 SCC has two kinds of probes- one is the main probe and other is the bore scope. The main probe is meant to ready the intensity of current, whereas the bore scope is means to see the beam spot on a ZnS Screen.
The main probe moves over the central spiral line of the hill whereas the bore probe moves across another hill after entering from its corner (Fig. 4).
To know whether the beam was shifting towards a particular area or not, need of a third probe was felt. For this reason the deflector was dismantled (to gain access to a port) and another probe called the deflector probe was installed (Fig. 4) along 25° azimuth.
Scheme of measurement
The beam curvature is not circular but a scalloped one. The local radius of curvature will thus change as the beam proceeds. The curvature will be less at the centre of the hill and more at the valley. Moreover, the three probes are not symmetrically placed with rest respect to the orbit trajectory. So, any data that is taken for the purpose of analysis has to take into account the scalloping of the beam vis-a-vis the position of the 3 probes.

Figure 4: Median Plane top view with main probe, bore probe and deflector probe.
Profile of the Beam
Beam profile with all the 3 probes was measured separately. It was observed that although good amount of beam was available up to the extraction radius ~ 670 mm in main probe and bore probe, but in deflector probe the beam current was falling at ~650 mm without application of any external 1st harmonic field (Fig. 5).

Figure 5: Beam profile on 3 probes without any external 1st harmonic.
This clearly indicated that beam was shifted towards main probe and bore probe region. Attempts were made to equalize the beam current in all the 3 probes by using different combinations of 1st harmonic amplitude (up to 40 gauss) and phase (1-360°). One of those several iteration has been illustrated in figure 6.
However, the distribution could not be equalized in all the three probes. From the shadowing of radial distribution of beam by using the said three probes, centering behaviour of the beam was quantified [2].

Figure 6: Beam profile on 3 probes with 15 Gauss 1st harmonic (phase 151°)
MAGNETIC FIELD MAPPING
These observations pointed out towards significant amount of imperfection in field and also towards the fact that this imperfection cannot be taken care of by the beam controlling parameters. Also the absolute phase measurement was done [3], which also pointed out towards the field imperfection. The remapping of the magnetic field was thus contemplated.
The magnetic field mapping was done by a search coil. The Calibration factor of the search coil was found out by the use of two NMR probes.The search coil was installed on a zig that was free to move in the horizontal plane. At any azimuth the search coil moved from -1.24inch to 26.4 inches. The mapping was done for different excitations of main magnet coils namely alpha and beta coils
The magnetic field mapping found that the average field near the central region was less by around 250 Gauss and the 1st harmonic amplitude was quite high ~ 45 Gauss near the extraction region.

Figure 7: 1st Harmonic field amplitude versus Radius

Figure 8: Average field versus Radius
In figure 7 and 8, the results of the magnetic field mapping is compared with that of the field mapping that was done earlier in year 2006.
REMARKS
The experimental observation that pointed out towards magnetic field imperfection was verified by remapping of the magnetic field. The correction of magnetic field is currently being done.
ACKNOWLEDGMENT
The authors would like to thank all persons who have contributed directly or indirectly, especially the control room personnel who have toiled through the day and the night in the endeavour towards a common goal.
 

What Will You Get?

We provide professional writing services to help you score straight A’s by submitting custom written assignments that mirror your guidelines.

Premium Quality

Get result-oriented writing and never worry about grades anymore. We follow the highest quality standards to make sure that you get perfect assignments.

Experienced Writers

Our writers have experience in dealing with papers of every educational level. You can surely rely on the expertise of our qualified professionals.

On-Time Delivery

Your deadline is our threshold for success and we take it very seriously. We make sure you receive your papers before your predefined time.

24/7 Customer Support

Someone from our customer support team is always here to respond to your questions. So, hit us up if you have got any ambiguity or concern.

Complete Confidentiality

Sit back and relax while we help you out with writing your papers. We have an ultimate policy for keeping your personal and order-related details a secret.

Authentic Sources

We assure you that your document will be thoroughly checked for plagiarism and grammatical errors as we use highly authentic and licit sources.

Moneyback Guarantee

Still reluctant about placing an order? Our 100% Moneyback Guarantee backs you up on rare occasions where you aren’t satisfied with the writing.

Order Tracking

You don’t have to wait for an update for hours; you can track the progress of your order any time you want. We share the status after each step.

image

Areas of Expertise

Although you can leverage our expertise for any writing task, we have a knack for creating flawless papers for the following document types.

Areas of Expertise

Although you can leverage our expertise for any writing task, we have a knack for creating flawless papers for the following document types.

image

Trusted Partner of 9650+ Students for Writing

From brainstorming your paper's outline to perfecting its grammar, we perform every step carefully to make your paper worthy of A grade.

Preferred Writer

Hire your preferred writer anytime. Simply specify if you want your preferred expert to write your paper and we’ll make that happen.

Grammar Check Report

Get an elaborate and authentic grammar check report with your work to have the grammar goodness sealed in your document.

One Page Summary

You can purchase this feature if you want our writers to sum up your paper in the form of a concise and well-articulated summary.

Plagiarism Report

You don’t have to worry about plagiarism anymore. Get a plagiarism report to certify the uniqueness of your work.

Free Features $66FREE

  • Most Qualified Writer $10FREE
  • Plagiarism Scan Report $10FREE
  • Unlimited Revisions $08FREE
  • Paper Formatting $05FREE
  • Cover Page $05FREE
  • Referencing & Bibliography $10FREE
  • Dedicated User Area $08FREE
  • 24/7 Order Tracking $05FREE
  • Periodic Email Alerts $05FREE
image

Our Services

Join us for the best experience while seeking writing assistance in your college life. A good grade is all you need to boost up your academic excellence and we are all about it.

  • On-time Delivery
  • 24/7 Order Tracking
  • Access to Authentic Sources
Academic Writing

We create perfect papers according to the guidelines.

Professional Editing

We seamlessly edit out errors from your papers.

Thorough Proofreading

We thoroughly read your final draft to identify errors.

image

Delegate Your Challenging Writing Tasks to Experienced Professionals

Work with ultimate peace of mind because we ensure that your academic work is our responsibility and your grades are a top concern for us!

Check Out Our Sample Work

Dedication. Quality. Commitment. Punctuality

Categories
All samples
Essay (any type)
Essay (any type)
The Value of a Nursing Degree
Undergrad. (yrs 3-4)
Nursing
2
View this sample

It May Not Be Much, but It’s Honest Work!

Here is what we have achieved so far. These numbers are evidence that we go the extra mile to make your college journey successful.

0+

Happy Clients

0+

Words Written This Week

0+

Ongoing Orders

0%

Customer Satisfaction Rate
image

Process as Fine as Brewed Coffee

We have the most intuitive and minimalistic process so that you can easily place an order. Just follow a few steps to unlock success.

See How We Helped 9000+ Students Achieve Success

image

We Analyze Your Problem and Offer Customized Writing

We understand your guidelines first before delivering any writing service. You can discuss your writing needs and we will have them evaluated by our dedicated team.

  • Clear elicitation of your requirements.
  • Customized writing as per your needs.

We Mirror Your Guidelines to Deliver Quality Services

We write your papers in a standardized way. We complete your work in such a way that it turns out to be a perfect description of your guidelines.

  • Proactive analysis of your writing.
  • Active communication to understand requirements.
image
image

We Handle Your Writing Tasks to Ensure Excellent Grades

We promise you excellent grades and academic excellence that you always longed for. Our writers stay in touch with you via email.

  • Thorough research and analysis for every order.
  • Deliverance of reliable writing service to improve your grades.
Place an Order Start Chat Now
image

Order your essay today and save 30% with the discount code Happy